Amprobe DMIII-Multitest Power-Quality-Recorder Manual de usuario Pagina 25

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 24
AMPROBE
DMIII MULTITEST
EN-23
6.2.2. Results of "MAN" mode
At the end of the test if
the insulation
resistance is lower
than R
MAX
(see Table2)
and the instrument
generates the
Nominal test Voltage,
the instrument emits a
double beep,
indicating the positive
outcome of the test
and displays a screen
similar to the screen to
the right.
M 05.06.01
Insulation Resistance
1.07 M
514V 15s
MAN 500V
Voltage during the Test
Duration of the Test
Test mode
Test voltage set
FUNC
VNOM
In order to evaluate the test you must compare the result with the limits indicated in the
Guidelines (see Table1).
The displayed result can be stored pressing the SAVE key twice (according to
paragraph 9.1).
If the Insulation
resistance is higher
than R
MAX
(see Table2),
the instrument emits a
double beep at the end
of the test indicating
the positive outcome
of the test and displays
one screen similar to the
screen to the right.
M 05.06.01
Maximum resistance value
which can be measured
(999 is displayed if a
rated voltage of 500V was
selected, see Table2).
> 999 M
523V 15s
MAN 500V
The symbol ">" means that
the resistance value is
higher than R
MAX
.
Test duration
FUNC
VNOM
The displayed result can be stored by pressing the SAVE key twice (refer to
paragraph 9.1).