DMIII MULTITEST
EN-25
6.2.4. "MAN" and "TIMER" mode faulty cases
If the instrument
detects the External
Power supply adapter
the instrument will
show the message
displayed to the right.
-.- -
M
---V 15s
REMOVE POWER
MAN 500V
Disconnect the External
Power Supply Adapter
If the instrument
detects a Voltage
between the input
terminals higher than
15V, the instrument
does not perform the
test and displays the
screen to the right for
5 seconds.
-.-
- M
---V 15s
VOLT IN INPUT
MAN 500V
ATTENTION: the test can't be
executed. Check that the circuit is
not energized.
This result can't be saved
If the instrument can't
generate the Nominal
Test Voltage it will
emit a long beep and
displays a screen
similar to the screen
to the right.
1.17
M
107V 15s
MAN 500V
ATTENTION: the test of
resistance R
ISO
was taken at a
voltage value lower than the
set rated voltage. Low
insulation case. This case
occurs under low insulation
conditions or in the presence of
capacitance on the installation.