GP-2 GeoTest
EN - 14
NOTE! A test with a result over 700Ω is not inserted in the calculation of average value.
Example:
1
st
measurement
Main display: 1,07Ω
Secondary display on the left-hand side: 1
Secondary display on the right-hand side: 1,07Ω
2
nd
measurement
Main display: 4,15Ω
Secondary display on the left-hand side: 2
Secondary display on the right-hand side: 2,61Ω
3
rd
measurement (not inserted in the average value)
Main display: 1018Ω
Secondary display on the left-hand side: 2
Secondary display on the right-hand side: 2,61Ω
Press CLR if you want to cancel the average resistance value and the no. of
measurements, which are included in the calculation (displayed on the
secondary displays on the right-hand side and on the left-hand side
respectively). A screen similar to the following will be displayed:
0.960.96 ΩΩ
- - - - - - - - -- - -
NOTE! The resistance measurement is taken according to a 4-wire voltmetric method.
Therefore, it is not affected by the resistance value of the cable used: it is not
necessary to calibrate the cables or their extension.
The test results can be stored in memory by pressing SAVE (see paragraph
5.1)
measured