1Users Manual• Mode d’emploi• Bedienungshandbuch• Manual d’Uso• Manual de usoACD-50NAV600A AC Navigator ClampACD-51NAV600A AC TRMS Navigator Clam
4Auto Power OffAfter idle 15 min The meter can work again by turning it on from the OFF position.Auto Power Off (APO) disable : Press “Downward” of Na
5Measuring VoltageVWARNING To avoid electrical shock, hazard or damage to meter, do not attempt to measurement that might exceed 1000 V DC or AC RMS
6Measuring CurrentOKII+(-I)=0OKOKIIII- Don’t clamp on any conductor while the meter power on.- ACD-50NAV, ACD-51NAV or ACD-53NAV has only AC current m
7AUTO SENSE mode:Display measurement result at AC only with RMS value or DC value, it depends on whichever is greater.AC mode: AC only with RMS value.
8In PEAK HOLD mode, the meter is activated to save the positive peak value and negative peak value. Positive peak value is displayed in PEAK MAX mode.
9DC A ZERO (For ACDC-52NAV, ACDC-54NAV Only)Remove the jaw out of the conductor. Press HOLD Key > 2 seconds to compensate the residual magnetism.HO
10"MAXMIN""MAXMIN""MAXMIN""MAXMIN"ENTER MAXMIN >2secNote :• Press HOLD key in MAX MIN mode to make the met
11Individual Harmonic Measurement (AC mode only)Select the “ “ indicator then press the navigator key to enter the individual harmonic mode. To quit f
12Measuring Active Power (W) / Power factor (PF)1. Single Phase Power MeasurementStep1. Set the rotary switch to the “W” position.Step2. Connect the R
132.Three Phase Power MeasurementA. 3 phase 3 wire balanced / unbalancedStep 1. Set the rotary switch to the “W” positionStep 2. Using the Navigator k
14Phase RotationLine 3Line 2Line 1Note :• Connect three phase of power source as shown above.• The test is only available while the system frequency
15c) When displayed “- - - - “ means it is unable to determine line phases.d) If it displays “LoU”, it is possible that you remove the test leads befo
16Measuring CapacitanceSet the rotary switch to the ““ position.CapacitanceCAUTION To avoid possible damage to the meter or to the equipment under t
17Measuring Temperature °C / °F ( For ACD-51NAV and ACDC-54NAV Only)Set the rotary switch to the “ ” position."°F"ENTERDon’t take any high
18Measuring μA Current (For ACD-51NAV Only)Set the rotary switch to the μA position.Disconnect
19Other Function :AUTO/MANUAL RangeSelect the “RANGE“ indicator, then press the navigator key to enter the manual range mode. To return to the auto ra
20VoltSenseThe red diamond shape of LED will illuminate, if there has electric field been detected from the jaw.BuzzerThe Meter beeps once for every va
21SPECIFICATIONS General Specications Display count : 10000 or 4000Measuring rate : 3 times / per secondOverrange display : “OL” or “-OL” .Auto Power
22Electrical SpecicationsAccuracy is ±(% reading + number of digits) at 23°C ± 5°C < 80%RH.(1) VoltageACD-50NAV, ACD-51NAV, ACDC-52NAV, ACD-53NAV,
23ACDC-52NAV, ACDC-54NAVFunction Range AccuracyDC A99.99A ± (1.5% + 0.2A)599.9A / 999.9A* ± (1.5% + 5dgt) **AC A0.10A ~ 99.99A50 ~ 60Hz ± (1.5% + 5dgt
ACD-50NAV600A AC Navigator ClampACD-51NAV600A AC TRMS Navigator ClampACDC-52NAV600A AC/DC TRMS Navigator ClampACD-53NAV1000A AC TRMS Navigator ClampAC
24(4) FrequencyACD-50NAV, ACD-51NAV, ACDC-52NAV, ACD-53NAV, ACDC-54NAVFunction Range AccuracyFrequency20.00 ~ 99.99Hz± (0.5% + 3dgt)20.0 ~ 999.9Hz0.02
25Overload protection: 1000 Vrms600 Arms For ACD-50NAV, ACD-51NAV, ACDC-52NAV1000 Arms For ACD-53NAV, ACDC-54NAVAccuracy dened for:Sine wave, ACA≧10A
26(9) Resistance & Continuity & Diode:ACD-50NAV, ACD-51NAV, ACDC-52NAV, ACD-53NAV, ACDC-54NAVFunction Range AccuracyResistance999.9 ± (1.0%
27(12) DC μA :ACD-51NAVFunction Range AccuracyDCμA 999.9 μADC ± (1.7% + 2dgt)**<1000dgt, add 3 dgt to the accuracy.Overload protection : 1000 VrmsM
28CAUTION Remove test leads from Meter before opening the battery cover or Meter case.
Limited Warranty and Limitation of Liability Your Amprobe product will be free from defects in material and workmanship for 1 year from the date of pu
ACD-50 Series CAT IV Navigator Clamps Series1 Jaw2 Voltsense Light3 Trigger4 Hold / DCA Zero Key5 Rotary Switch6 Display7 Navigator Key8 V / / Inp
CONTENTS SYMBOLS ... 1UNPACKING AND INSPECTION ..
1SYMBOLSCaution ! Risk of electric shockCaution ! Refer to the explanation in this ManualApplication around and removal from hazardous live conduct
2Measurement category:V : Category III 1000 Volts AC & DC, and Category IV 600 Volts AC & DC.A : Category III 1000 Volts AC & DC, and Cate
3o DC Current- Up to 600A (ACDC-52NAV)- Up to 1000A (ACDC-54NAV)o Capacitanceo Temperature Measurement in °C / °F (ACD-51NAV, ACDC-54NAV only) - Temp.