CURRENT (1ms PEAK HOLD)
Specified accuracy +/-40 digits for changes > 1 ms in duration.
Range Resolution Accuracy
400A 0.1A ±(2%rdg+43dgt)
1000A 1A ±(2%rdg+43dgt)
FREQUENCYCOUNTERSENSITIVITY
INPUTRANGE
MINIMUM SENSITIVITY
(RMSSINEWAVE)
400A 3A
1000A 30A
400V 3V
1000V 30V
(Maximum input for
specified accuracy
=10 x range or1000V)
40 Hz-2 kHz
FREQUENCY(AC coupling)
Range Resolution Accuracy Minimum Input Frequency
100Hz 0.01Hz
1kHz 0.1Hz
• Overload protection: 1000Vrms AC;< 1000000 VxHz
±(0.2%rdg+4dgt) 10Hz
DIODECHECK
Range Resolution Accuracy Test Current Test Voltage
Diode 1mV ±(1.0%rdg+2dgt) approx. 1.65 mA <3.3V
• Overload protection: 600VRMS
RESISTANCE
Range Resolution Accuracy Max Test Voltage Overload Protection
400Ω 0.1Ω 3.3V
4kΩ 1Ω ±(1%rdg+3dgt)
1.25V
600Vrms
• Instant continuity mode, built-in buzzer sounds when resistance is less than 10.0Ω.
built-in buzzer
sounds when
reading is
below approx.
100mV
AUDIBLE CONTINUITYTEST
Range Resolution Accuracy Test Current Test Voltage
approx. 1.65 mA <3.3V
• Overload protection: 600VRMS
1mVDiode