Amprobe TM45A Especificaciones Pagina 31

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 52
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 30
2928
For detailed specifications go to www.Amprobe.eu
DGC-1000A Earth Resistance Clamp Meter
B TRMS current measurement including
leakage current
B Earth loop resistance range 0.025 to 1500 Ω
B Data logging capability - 116 records
B Individual data point recording
B User selectable High & Low ohm
audible alarms
B Simple and easy to use, no test
leads or auxiliary electrodes
necessary
B Jaw opening 23 mm
B Safety acc. EN 61010, CAT III/300V,
CAT II/600 V
Earth Resistance Testers
Ensure safe operation and reduce power quality problems with Amprobe earth resistance
testers. Use the DGC-1000A for earth loop resistance checks.
TELARIS EARTH-TEST
Earth Resistance Tester
B Earth resistance measurement
B Specific earth resistance measurement
B The integrated constant current measurement
principle allows the application in sandy as well
in rocky environments
B Monitoring and display of auxiliary earth and
probe resistances
B Displays of the actual test current
B Automatic and manual and frequency selection
for noise reduction
B Test voltage preselection
B Clear and large LCD providing the user
with an optimum visual indication of
both test values and limits
B Auto power off
B Earth resistance measurement 2/3 and
4-pole
B Specific earth resistance measurement in
compliance with the Wenner principle
Software for Reporting of Installation and Machinery Tests
-
Comfortable, extensible software for administration and reporting of installation and machinery tests
-
Clear and logic hierarchy
-
Directory structure, easy to handle (customer, site, department, UUT, Test), is always visible
-
Measurement data transfer from test instrument to PC
-
Print of test reports
-
Export function for listings in text or Excel format
-
New data records for UUT will be automatically created. Test results can be printed out quickly and easily
Special Features (es control PROF only)
-
Administration of user authorisation for four different user groups
-
Comparision and update of different es control databases
-
Evaluation of test result for installation test
Software in three modules avaliable:
B Module „es control INST“ for reporting and documentation of installation tests in compliance with EN 60364,
DIN VDE 0100/0105, ÖVE E 8001, NIN/NIV, 16th Edition
B Module „es control PAT“ for reporting and documentation of machinery and appliance tests in compliance with
EN 60204, DIN VDE 0701/0702, ÖVE E8701
B Complete package „es control COMPL“ for reporting and documentation of installation, machinery and appliance
tests
Free demo version available
The software es control support the following products:
Genius 5035
TELARIS 0100-PLUS
TELARIS RCD-TEST
• TELARIS ISO-PRO
• TELARIS LOOP-TEST
ALLTEST
es control, es control PROF (professional)
FEATURES
Earth Resistance
AC Current
DGC-1000A
0.025Ω - 1500Ω
0.2A - 30.00A
For detailed specifications go to www.Amprobe.eu
EARTH RESISTANCE & MACHINERY TEST
ALLTEST
Machinery Tester acc. EN 60204-1
B Insulation measurement 500 V DC
B Continuity testing (10 A AC), resistance and
voltage drop
B Continuity testing, 6 V AC/10 - 25 A AC
B Withstanding voltage to 4000 V
B Leakage current 0 - 50 mA
B Residual voltage measurement
B PC interface (RS-232) + protocol software
B Specification of limit value
B Acoustic signal for „OK“, „Fault“ and
„Warning“
B 700 memory locations
B Rugged waterproof case
B Safety acc. EN 61010, CAT III/600 V
Machinery Tester
FEATURES TELARIS EARTH-TEST
Display 3 digit; LCD; 1999 Digits
Meas. principle
Meas. range /
Resolution
Current-voltage measurement
Earth resistance
0.05...19.99 Ω / 0.01 Ω
20...199.9 Ω / 0.1 Ω
200...1999 Ω / 1 Ω
Tolerance
±(4% rdg.+2 Digits / 0.1Ω)
Probe/aux. resistance
0.1...50 kΩ / 0.1 kΩ
Tolerance ±(10% rdg.+3Digits)
Test current display 0.1...13 mA / 0.1 mA
Tolerance ±(10% rdg.+3 Digits)
Max. test current approx. 12 mA
Technical data
according to EN 60204-1
Range
(Resolution)
Test
voltage
Test
current
Test time comments
Withstanding +
burn funcion (BURN)
1kV
0-4kV
1000V AC
0-4000V AC
0,5-500mA
1s to 10min
Display of test voltage
and leakage current,
power 500 V A
Leakage current
0-50mA
(0,1mA)
1s to 10min for VDE 0702
Insulation
0-99,99MΩ
(10kOhm)
500V DC 2,2mA 1s to 10min
Possibility to adjust
a limit value
Continuity
Voltage drop
measurement
0-1 Ω
(0,001 Ω)
0-10V
(0,01V)
12V AC or 6V
AC
6 V AC
>10A or
>10<25A
AC*
1s to 15s
For testing the voltage
drop, the wire gauge ist
selectable:
1-1,5-2,5-4 and >6mm2
Residual voltage measurement
internal and external
0-10,0s limit value 60V
limit value 1 s resp. 5 s
adjustable
FEATURES TELARIS EARTH-TEST
Test voltage max. 25 V / 50 V AC selectable
Frequency 127 Hz / 140 Hz selectable
Protection class II (double insulated)
Safety complying with IEC61010/EN61010/DIN VDE 0411
IEC61557/EN61557/DIN VDE 0413
Measurement category CAT III / 300 V
Pollution degree 2
Protection degree IP 50
Auto-Power-Off after approx. 5 min.
Power supply 6 x 1.5 V, IEC LR6
Memory 700 memory locations
Case rugged weatherproof case
Power supply External power supply 230 V AC / 50 Hz
Standard Accessories Cable for measurment of voltage stability, 1 pistol and 1 alligator clip, cable for insulation
measurement, cable with 4 alligator clips for low ohms measurement, protokol software and
interface cable to connect ALLTEST and PC, calibration certificate
Vista de pagina 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 51 52

Comentarios a estos manuales

Sin comentarios