3
Specifications
GENERAL SPECIFICATIONS
Display: LCD with a max. reading of 4000
Range control: Auto range & Manual range control
Polarity: Automatic negative polarity indication
Zero adjustment: Automatic
Overrange indication: The “OL” or “-OL” display
Low battery indication: Display “
“ sign
Data hold: Display “
” sign
Relative measurement: Display “
” sign
Auto Power Off: Display “
” sign. After 10 minutes without switching modes
or pressing a key the meter will switch to standby mode. Press
any key or switch selector switch to exit standby mode. To dis-
able auto power off, press and hold SELECT key while rotating
the selector switch from off position.
Safety standards:
EMC/LVD. The meter is up to the standards of
EN61010-1:2001; Double Insulation, Pollution Degree 2,
Overvoltage Category III-600V.
Operating environment: Temperature 0°C to 40°C (32°F to104°F)
Humidity °< 80% RH
Storage environment: Temperature -20°C to 60°C (-4°F to140°F)
Humidity °< 90% RH
Power supply: 2x1.5V “AA” batteries
Dimension: 156(H) x 86(W) x 38(D) mm (6.1“ x 3.4“ x 1.5“)
Weight: Approx. 260g (9.2 oz.) including battery
ELECTRICAL SPECIFICATIONS
Accuracies are ±(% of reading + number in last digit) at 23 ± 5°C, °< 75% RH
DC VOLTAGE
Range Accuracy Resolution
400mV
± ( 0.5%+2 )
0.1mV
4V 1mV
40V 10mV
400V 100mV
600V ± ( 0.8%+3 ) 1V
Overload protection: 600V DC or AC RMS
Impedance: 10MW, More than 100MW on 400mV range
Warning! Dangerous Voltage (Risk of electric shock).
Caution! Refer to the user’s manual before using this Meter.
Double Insulation (Protection Class II).
Alternating Current (AC).
Direct Current (DC).
Complies with European Directives
Either DC or AC.
Ground (maximum permitted voltage between terminal and ground).
The RESPONSIBLE BODY shall be made aware that, if the instrument
is used in a manner not specified by the manufacturer, the protection
provided by the instrument may be impaired.
The finger or any part of your body shall not be beyond the barrier of the
test probe when measuring.
Conforms to relevant Australian standards
Do not dispose of this product as unsorted municipal waste