8
MAINTENANCE
The meter is calibrated in house before shipping. When properly maintained, the meter will
maintain its accuracy as specified.
Cleaning
Use a damp cloth and mild soap to clean the case of the MAN15 Manometer, do not use harsh
detergents or abrasives as these may mar the finish or damage the unit’s case with an adverse
chemical reaction.
SPECIFICATIONS
Operating Conditions
Compensated temperature range: 0~50 °C (32~122 °F)•
Operating temperature range: 0~50 °C (32~122 °F)•
Storage temperature range: -20~60 °C (-4~140 °F)•
Operating Humidity Max. 80% RH•
Power Supply : 9 volt battery
Exceeding Maximum pressure will cause permanent sensor damage.
Range: 0 ~ ±2 psi (MAN02)
0 ~ ±15 psi (MAN15)
0 ~ ±30 psi (MAN30)
Power Supply :
Conversion & Resolution
1 psi = Resolution
inch of H₂O
0.401 0.01
psi 0.0145 0.001
mbar 1 0.1
bar 0.001 0.001
mm of H₂O
10.2 1
1 psi = Resolution
bar 0.068948 0.001
mmHg 51.715073 0.5
ozin2 16.000844 0.2
Kgcm2 0.070309 0.001
psi 1 0.01
inH2O 27.679904 0.3
kpa 6.894757 0.1
ftH2O 2.306659 0.02
inHg 2.036021 0.01
cmH2O 70.306955 1
mbar 68.94757 1
Accuracy: ±0.3% of full scale at ±25°C (77°F)
Repeatability: ±0.2%~0.5% of full scale
Linearity / Hysteresis: ±0.29% to 1% of full scale
Combined accuracy: ±1.0% of full scale
Dimension: 72 x 182 x 30mm
Unit Weight: Approx. 220 (o.485lb) (with battery)
Response Time: 0.5 seconds
Format: Baud rate: 2400 bit/sec
Data bit: 8, Stop bit: 1
Parity: none