Amprobe Multitest-1000 Continuity-Tester Manual de usuario Pagina 17

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 122
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 16
4.1.3. Anomalous cases during "AUTO", "R+TIMER", "R-TIMER" tests
)
CAL LOW
5.75
216 mA
5 s
AUTO
In case a value of Ravg or R+ or R-
greater than or equal to 5 but less
than 99.9 was detected, at the end of
the test the instrument emits a long
beep and displays a screen similar to
the screen alongside.
Only in the case
R+TIMER or R-
TIMER was
selected.
ATTENTION: value of R
avg
g
reater than 5
AUTO mode.
)
CAL LOW
>
99.9
- - - mA
AUTO
In AUTO mode, if a Ravg or R+ or R-
greater than 99.9 was detected, at
the end of the test the instrument emits
a long beep and displays the screen
alongside.
99.9 is the
maximum value
that can be
measured in the
LOW AUTO or
R+ or R-mode.
ATTENTION: value of R
avg
or R+ or R- is too
hi
g
h.
AUTO mode.
)
CAL LOW
>
99.9
- - - mA 4 s
R+
TIMER
In the case mode R+TIMER or R-
TIMER was selected and a R+ or R-
greater than 99.9 was detected, the
instrument emits an intermittent beep
during the test, a long beep at the
end of the test and displays the screen
alongside.
99.9 is the
maximum value.
ATTENTION:
value of R+ or R-
is too high.
)
CAL LOW
ν
0.00
219 mA
A
UTO
In case that:
R
MEASURED
-R
CALIBRATION
< 0 the
instrument displays the screen
alongside.
Blinking CAL.
ATTENTION: R
MEASURED
-R
CALIBRATION
< 0
AUTO mode.
SAVE
SAVING
: The tests can be stored pressing the SAVE key twice (refer
to paragraph 5.1).
Vista de pagina 16